| Into | Inkcazo | Umgangatho |
| I-voltage yetyhubhu ye-x-ray eqhelekileyo | 160kV | I-IEC 60614-2010 |
| I-voltage yetyhubhu yokusebenza | 40~160KV | |
| Ityhubhu ephezulu yangoku | 1.5mA | |
| Izinga eliphezulu lokupholisa eliqhubekayo | 240W | |
| Ubuninzi be-filament current | 3.5A | |
| Umbane ophezulu wefilament | 3.7V | |
| Izinto ekujoliswe kuzo | I-Tungsten | |
| I-engile ekujoliswe kuyo | 25° | I-IEC 60788-2004 |
| Ubungakanani bendawo egxile kwifokha | 0.8x0.8mm | IEC60336-2005 |
| I-engile yokugubungela umqadi we-X-ray | 110°x20° | |
| Ukuhluzwa okungokwemvelo | 0.8mmBe kunye ne0.7mmAl | |
| Indlela yokupholisa | Ioyile efakwe emanzini (70°C Ubuninzi.) kunye neoyile yokupholisa nge-convection | |
| Ubunzima | 1020g |
Funda izilumkiso ngaphambi kokusebenzisa ityhubhu
Ityhubhu ye-X-reyi iya kukhupha i-X-reyi xa inikwe amandla nge-voltage ephezulu, kufuneka ulwazi olukhethekileyo kwaye kufuneka kuthathwe amanyathelo okulumka xa iphathwa.
1. Yingcali eqeqeshiweyo kuphela enolwazi ngetyhubhu ye-X-Ray ekufuneka ihlanganise, igcine kwaye isuse ityhubhu.
2. Kufuneka kuthathwe unonophelo olwaneleyo ukuze kuthintelwe ukungqubana okukhulu kunye nokungcangcazela kwityhubhu kuba yenziwe ngeglasi ebuthathaka.
3. Ukukhuselwa kwemitha yeyunithi yetyhubhu kufuneka kuthathwe ngokwaneleyo.
4. Ityhubhu ye-X-ray mayiphathwe ngokucoca, ukomisa ngaphambi kokuba ifakelwe. Kufuneka uqinisekise ukuba amandla okukhusela ioyile angaphantsi kwe-35kv / 2.5mm.
5. Xa ityhubhu ye-x-ray isebenza, ubushushu beoyile akufuneki bube ngaphezulu kwama-70°C.
Ubungakanani obuncinci be-oda: 1pc
Ixabiso: Ingxoxo
Iinkcukacha zokuPakisha: 100pcs ngebhokisi nganye okanye eyenziwe ngokwezifiso ngokwenani
Ixesha lokuHanjiswa: Iiveki ezi-1 ~ 2 ngokobungakanani
Imigaqo yokuhlawula: 100% T/T kwangaphambili okanye i-WESTERN UNION
Amandla okubonelela: 1000pcs/ngenyanga