I-China X-RAY TUBE EFANELEKILEYO TOSHIBA E7242 Umvelisi kunye noMboneleli | Sailray
I-X-RAY TUBE EFANELEKILEYO TOSHIBA E7242

I-X-RAY TUBE EFANELEKILEYO TOSHIBA E7242

I-X-RAY TUBE EFANELEKILEYO TOSHIBA E7242

Inkcazelo emfutshane:

Isicelo: Indibano yetyhubhu ye-X-ray kuzo zonke iimviwo zesiqhelo zokuxilonga kunye nesiqhelo
okanye iindawo zokusebenzela zeradiographic kunye nefluoroscopic
◆ Iimpawu zofakelo : 12.5° Rhenium-Tungsten molybdenum target (RTM)
◆ Iindawo ekugxilwe kuzo: Encinci 0.6, Enkulu: 1.2
◆ Ubona buninzi bombane wetyhubhu: 125kV
◆Ihlala ne-IEC60526 uhlobo lwezixhobo zentambo ze-high-voltage
◆ Ijeneretha yamandla ombane aphezulu kufuneka ihambelane ne-IEC60601-2-7
◆ UHlelo lwe-IEC (IEC 60601-1:2005): Udidi lwe-I ME IZIXHOBO

Iinkcukacha zeMveliso

Intlawulo kunye neMiqathango yokuThumela:

Iithegi zeMveliso

Isalathiso esisemgangathweni

Isalathiso esisemgangathweni

Izihloko

EN 60601-2-54:2009 Izixhobo zombane zonyango-Icandelo 2-54: Iimfuno ezizodwa zokhuseleko olusisiseko kunye nokusebenza okubalulekileyo kwezixhobo zeX-ray zeradiography kunye neradioscopy
IEC60526 Iplagi yentambo ene-voltage ephezulu kunye noqhagamshelo lwesokhethi yezixhobo zonyango zeX-reyi
IEC 60522:1999 Ukumiselwa kokucoca ngokusisigxina kweendibano zetyhubhu ze-X-ray
IEC 60613-2010 Umbane, i-thermal kunye neempawu zokulayishwa kweetyhubhu ze-X-ray ze-anode ezijikelezayo zokuxilongwa kwezonyango
IEC60601-1:2006 Izixhobo zombane zonyango-Icandelo 1: Iimfuno ngokubanzi zokhuseleko olusisiseko kunye nokusebenza okubalulekileyo
IEC 60601-1-3:2008 Izixhobo zombane zonyango-Icandelo 1-3: Iimfuno ngokubanzi zokhuseleko olusisiseko kunye nokusebenza okubalulekileyo
IEC60601-2-28:2010 Izixhobo zombane zonyango-Icandelo 2-28: Iimfuno ezizodwa zokhuseleko olusisiseko kunye nokusebenza okubalulekileyo kweendibano zetyhubhu zeX-ray zokuxilongwa kwezonyango.
IEC 60336-2005 Izixhobo zombane zonyango-iindibano zetyhubhu zeX-ray zokuxilongwa kwezonyango-Iimpawu zeendawo ezigxininisekileyo

Inkcazo

● Uluhlu luqulunqwe ngolu hlobo lulandelayo:

MWHX7110A

Umbhobho

A

Isokhethi yamandla ombane aphezulu ene-90 degree direction

MWTX71-0.6/1.2-125

B

Isokethi yamandla ombane aphezulu ene-270 degree direction

Idatha yoBugcisa

Ipropati

Inkcazo

Umgangatho

Amandla engeniso yesiqhelo ye anode

F 1

F 2

IEC 60613

20kW(50/60Hz)

40kW(50/60Hz)

 

Umthamo wokugcina ubushushu be-Anode

110 kJ ( 150kHU)

IEC 60613

Ubuninzi bomthamo wokupholisa we-anode

500W

 
Umthamo wokugcina ubushushu

900kJ

 
Max. ukutshatyalaliswa kobushushu okuqhubekayo ngaphandle kwe-Air-circular

180W

 
Izinto ze-AnodeI-Anode yokwambathisa izinto eziphezulu

I-Rhenium-Tungsten-TZM(RTM)

I-Rhenium-Tungsten-(RT)

 
I-engile ekujoliswe kuyo (Ref: i-axis yesalathiso)

12.5 °

IEC 60788

X-reyi indibano yetyhubhu yokucoca ngokwendalo

1.5 mm Al / 75kV

IEC 60601-1-3

Indawo ekujongwe kuyo ixabiso legama (ama)

F1(ingqwalasela encinci)

F2(ujoliso olukhulu)

IEC 60336

0.6

1.2

 
I-X-ray ityhubhu yombane ngegamaRadiographicI-Fluoroscopic

125kV

100kV

IEC 60613

Idatha yokufudumeza kwe-cathode Max. yangoku

Ubuninzi bombane

≈ / AC, <20 kHz

 

F1

F 2

 

5.1A

≈79V

5.1 A

≈1214 V

 
Ukuvuza kwemitha kwi-150 kV / 3mA kumgama we-1m

0.5mGy/h

IEC60601-1-3

Ubuninzi bemitha yemitha

443 × 443mm kwi-SID 1m

 
Ubunzima bendibano yetyhubhu ye-X-ray

Malunga. 18kg

 

Iimeko zokusebenza, ukugcinwa kunye nokuthutha

Imida

Imida yoMsebenzi

Imida yezoThutho noGcino

Ubushushu obugciniwe

Ukusuka kwi-10ukuya ku40

Ukususela kwi-20to 70

Ubuntu bezizalwane

≤75%

≤93%

Uxinzelelo lweBarometric

Ukusuka kwi-70kPa ukuya kwi-106kPa

Ukusuka kwi-70kPa ukuya kwi-106kPa

 

Amaxabiso angundoqo eStator

I-stator yesigaba esi-1

Indawo yovavanyo

C-M

C-A

Ukuxhathisa komoya

≈18.0…22.0Ω

≈45.0…55.0Ω

Ubukhulu bombane obuvumelekileyo bokusebenza (ukuqhuba)

230V±10%

Cebisa ukusebenza kwamandla ombane (ukusebenza)

160V±10%

I-voltage ye-Braking

70VDC

I-voltage esebenzayo ikwi-exposure

80Vrms

I-voltage esebenzayo kwi-fluoroscopy

20V-40Vrms

Ixesha lokuqalisa (kuxhomekeke kwinkqubo yokuqalisa)

1.2s

Isilumkiso kuMsebenzi

1 .Ukusasazeka kweX-reyiukhuselo

Le mveliso izalisekisa iimfuno ze-IEC 60601-1-3.

Le ndibano yetyhubhu ye-X-reyi ikhupha imitha ye-X-reyi ekusebenzeni.Kuphela ngokuhambelanayo kuphela ngabasebenzi abaqeqeshiweyo nabaqeqeshiweyo abavumelekileyo ukuba basebenzise indibano yetyhubhu ye-X-ray.

Iziphumo ezifanelekileyo ze-physiologic zinokubangela ingozi kwisigulana, ukuveliswa kwenkqubo kufuneka kuthathe ukhuseleko olufanelekileyo ukunqanda imitha ye-ionization.

2.Dielectric 0il

Indibano yetyhubhu yeX-reyi ine-dielectric 0il equlathe uzinzo lombane ophezulu. Njengoko iyityhefu kwimpilo yabantu,ukuba ibonakaliswe kwindawo engathintelwanga,kufuneka ilahlwe ngokulandelayo kummiselo wendawo.

3 .I-Operation Atmosphere

Indibano yetyhubhu yeX-reyi ayivumelekanga ukuba isetyenziswe kwi-atmosfera yegesi enokutsha okanye edlayo ·

4.Lungisa iTubhu yangoku

Ngokuxhomekeke kwiimeko zokusebenza,iimpawu zomcu zinokutshintshwa.

Olu tshintsho lunokuthi lube sesichengeni esingaphezulu kweqondo lokuvezwa kwimibhobho yeX-reyi.

Ukuthintela indibano yetyhubhu yeX-reyi ukuba yonakale,lungisa i-tube yangoku rhoqo.

Ngaphandle xa ityhubhu ye-X-reyi ine-arcing Ingxaki kwi-alukusetyenziswa ixesha elide,ukulungiswa kombhobho wangoku kuyafuneka.

5.X-ray Tube Housing Ubushushu

Sukuchukumisa umphezulu wezindlu zetyhubhu yeX-ray nje emva kokusebenza ngenxa yobushushu obuphezulu.

Hlala ityhubhu yeX-reyi ukuze upholiswe.

6.Imida yokusebenza

Ngaphambi kokusetyenziswa,Nceda uqinisekise imeko yokusingqongileyo ingaphakathi kwiiIimits zokusebenza.

7.Nakuphi na ukungasebenzi kakuhle

Nceda uqhagamshelane ne-SAILRAY ngokukhawuleza,ukuba kukho ukungasebenzi kakuhle kwendibano yetyhubhu ye-X-ray kubonwa.

8.Ukulahlwa

Indibano yetyhubhu yeX-reyi kunye netyhubhu iqulethe izinto ezifana neoyile kunye nesinyithi esinzima apho ukulahla okusingqongileyo kunye nokufanelekileyo ngokuhambelana nemithetho esemthethweni yesizwe kufuneka kuqinisekiswe. Ukulahlwa njengenkunkuma yasekhaya okanye yemizi-mveliso akuvumelekanga. Umenzi unazo. ulwazi olufunekayo lobuchwephesha kwaye iya kuthatha indibano yetyhubhu yeX-reyi emva ukulahlwa.

Nceda uqhagamshelane nenkonzo yabathengi ngale njongo.

Iigophe zokukhutshwa kwecathode

图片

Ukuba (A) INdawo yokuJonga eNcinci

Imifanekiso ye-1.2

Ukuba (A) Indawo enkulu ekugxilwe kuyo

Ukulayisha okukodwa kunye nochungechunge

Iimeko:ISigaba seSithathu seTube Voltage

I-Stator Power Frequency 50Hz/60Hz

Ndim i)

umfanekiso we-4

T(s)

Ndim i)

Imifanekiso ye3

T(s)

Ukufudumeza kunye nokupholisa ijika le-anode

  IEC60613

图片5

Ukufudumeza kunye nokupholisa ijika lendibano yetyhubhu ye-X-reyi

Iimpawu zeThermal zezindlu

图片6

X-reyi indibano ityhubhu imizobo enomlinganiselo

SRMWHX7110A

Umfanekiso we7

INdibano yoHluzo kunye neCandelo lokuwela iZibuko

Umfanekiso we27

I-Rotor Connector Wiring

Umfanekiso we28

  • Ngaphambili:
  • Okulandelayo:

  • Ubuncinci boMyalelo woMyalelo: 1pc

    Xabiso:Negotiation

    Iinkcukacha zokupakisha: I-100pcs ngebhokisi okanye ilungiselelwe ngokobuninzi

    Ixesha lokuhambisa: Iiveki ze-1 ~ 2 ngokobuninzi

    Imigaqo yentlawulo: 100% T/T kwangaphambili okanye WESTERN UNION

    Ukubonelela ngeAmandla: 1000pcs / ngenyanga

    Bhala umyalezo wakho apha kwaye uwuthumele kuthi